Terminated

EliA RF IgM Well, REF 14-5600-01 Product Usage: Usage: EliA RF IgM is intended for the in vitro quantitative measurement of IgM class rheumatoid factor antibodies in human serum and plasma (Li-heparin and EDTA) to aid in the diagnosis of rheumatoid arthritis in conjunction with other laboratory and clinical findings. EliA RF IgM uses the EliA IgM method on the instrument Phadia 100, Phadia 250, Phadia 2500 and Phadia 5000.

FDA Recall Z-1907-2019 · Posted None

Recalling FirmPhadia Ab
Recall StatusTerminated
Product CodeDHR
Date Initiated2019-05-10
Date Terminated2020-06-02
Product Quantity30 kits

Reason for Recall

Potential for reporting low assay results

Firm Action

On May 10, 2019, Corporate Thermo Fisher Scientific issued URGENT FIELD SAFETY NOTICES to customers by FedEx. Actions To Be Taken: Please review patient results obtained from lot 0085 and make note of the following: 1. If you have any of the above referenced material on hand, please discontinue use immediately and discard per your internal procedures. Notify all affected users in your facility to do the same. 2. All samples reported as positive (> 5 IU/ml) are positive, however the true concentration could be higher than reported. 3. The true concentration of a reported sample equal to 3.5 5 IU/ml could potentially be a positive sample. 4. Negative results between 2 IU/ml and 3.5 IU/ml are potentially false negative results. 5. Negative results < 2 IU/ml could be higher, but would still be reported as a negative result with a 60% higher concentration. 6. Complete and return the Customer Response Form. If you have any questions concerning this issue, please contact Phadia US, Inc. Technical Support at 1-800-346-4364, option #2.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide n the states of NJ, NY, MI, WV, TX including PR

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.