Baxter EXACTAMIX Inlet, Non-Vented, High-Volume Inlet, REF H938173 Product Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding device used in the pharmacy to compound multiple source ingredients into one final solution.

FDA Recall Z-1908-2018 · Posted None

Recall StatusTerminated
Product CodeLHI
Date Initiated2018-04-25
Date Terminated2020-12-16
Product Quantity98700 units

Reason for Recall

Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiration date,

Firm Action

The firm initiated the recall by letter on 04/25/2018. The letter identified the affected product, problem and actions to be taken. the firm requested the return of the product. Distributors were directed to conduct a sub-recall, requesting their customers return the product to the entity from whom they received the product.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution - US Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.