Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

FDA Recall Z-1910-2019 · Posted 2019-07-01

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeLLZ
Date Initiated2019-05-16
Date Terminated2023-04-14
Product Quantity145 (87 US 103 OUS)

Reason for Recall

There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.

Firm Action

On May 20, 2019, the firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter date 5/20/2019 to their consignees. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: discontinue use of the UV study management functionality for study split/study info updates until a correction is available and to wait for a GE representative to contact them to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: FL, GA, IL, IN, LA, MA, MO, NJ, NY, OR, PA, SC, SD, TX, and WA; and countries of : Brazil, Canada, Chile, France, French Polynesia, Germany, India, Italy, Japan, Kuwait, Lebanon, Singapore, Spain and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.