ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)
FDA Recall Z-1910-2025 · Posted 2025-06-05
Advertisement
Recalling FirmHealthmark Industries Co., Inc.
Recall StatusOpen, Classified
Product CodeMNL
Date Initiated2025-05-02
Product Quantity193035 units
Reason for Recall
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
Firm Action
An "URGENT PRODUCT RECALL" notification letter was sent to customers beginning 5/2/25.
Recommended Mitigations/ User Recommendations
Please discontinue use of the Pre-Filled Water Syringes containing non-sterile water, which can be identified by a red cap. Any unused product can be returned for credit or replacement.
Customer are requested to take the following actions
1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined.
2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected products have been transferred.
3. Contact Customer Service via email at [email protected] or 800-521-6224 to return your unused affected product or if you have any questions.
4. Complete and return the response form Getinge by e-mailing a copy to [email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide and Internationally to countries of: Canada, Malaysia.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.