Terminated

Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.

FDA Recall Z-1912-2009 · Posted 2009-09-08

Recalling FirmVital Signs Inc
Recall StatusTerminated
Product CodeGAA
Date Initiated2009-06-30
Date Terminated2012-08-27
Product Quantity23,429 kits

Reason for Recall

Not compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module.

Firm Action

Vital Signs, Inc. issued an "Urgent Medical Device Correction" letter dated June 30, 2009 to all direct and indirect accounts by first class mail. Consignees were informed of the affected device, asked to complete a Confirmation form for return and to return unused product for replacement. For further questions, contact Vital Signs, Inc. at 1-973-790-1330 extension 404.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- US, Canada, United Arab Emirate, Venezuela, Guam, Korea and United Kingdom.