Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/15 cm, sterile, Boston Scientific, Spencer, IN; REF 26-226, UPN M001262260. The product is used with a radiofrequency generator for the thermal coagulation necrosis of soft tissues, including partial or complete ablation of nonresectable liver lesions.

FDA Recall Z-1913-2008 · Posted 2008-09-11

Recall StatusTerminated
Product CodeGEI
Date Initiated2008-04-30
Date Terminated2009-08-04
Product Quantity7,460

Reason for Recall

The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.

Firm Action

Consignees were notified by recall letter on 4/30/08, sent fexex priority mail, to cease use of the products and to return them to the firm. For assistance, contact Boston Scientific at 1-508-683-4186.

Root Cause

Device Design

Distribution Pattern

Nationwide, Argentina, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Iraq, Ireland, Italy, Jordan, Korea, Lebanon, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, and United Arab Emirates.