Triathlon AR 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-0-701, 6541-0-702, 6541-0-703, 6541-0-704, 6541-0-705, 6541-0-706, 6541-0-707, & 6541-0-708; Triathlon MIS Instruments; Non Sterile Stryker Orthopaedics 325 Corporate Drive Mahwah, NJ 07430 The Triathlon AR 4:1 Cutting Blocks and the Triathlon MIS 4:1 Cutting Blocks are used to prepare the Femur (bone cuts) for the Triathlon femoral component.
FDA Recall Z-1916-2009 · Posted 2009-08-26
Advertisement
Recalling FirmStryker Howmedica Osteonics Corp.
Recall StatusTerminated
Product CodeLXH
Date Initiated2007-02-12
Date Terminated2009-10-15
Product Quantity690 total units - both varieties
Reason for Recall
Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle.
Firm Action
Important Product Correction letters were sent to all branches/agencies on February 12, 2008 by Federal Express. The letter asked consignees to examine their inventory and hospital locations to identify the affected product. Customers were asked to perform a functional inspection by assembling the Triathlon Impactor/Extractor Handle to all of their MIS Cutting Blocks. If the cutting block does not assemble, customers are to list the affected lot codes on the Attached Product Accountability Form and fax a copy to (201) 831-6069. The letter states that product should not be returned until further instructions are received. Questions should be directed to (201) 831-5825.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.