Terminated

Automated External Defibrillator(AED) of brand name "Philips HeartStart HS1 OnSite" (Model number M5066A). Device is designed for use on a person in sudden cardiac arrest who is unresponsive when shaken and not breathing normally

FDA Recall Z-1916-2010 · Posted 2010-06-28

Recall StatusTerminated
Product CodeMKJ
Date Initiated2010-04-30
Date Terminated2012-12-10
Product Quantity205

Reason for Recall

Seventeen AEDs failed production line testing (Final Acceptance Test).

Firm Action

The firm notified U.S. distributors on 4/30/2010 via phone call and e-mail to obtain consignee location information. A website providing information was activated on 4/30/2010. On 5/3/2010 Urgent Medical Device Recall letters, dated April 29, 2010 were sent to direct customers in the US. The letters explained the reason for the recall and Customers were asked to verify that they have the product subject to recall. Replacement AEDs will be provided at no charge to the customer. Field Safety Notices were attached to the letters and provided a list of actions for customers to take. Customers should check the serial number on their devices against the list provided, contact Philips Healthcare to arrange for replacement, and ensure that necessary staff are aware of the recall. For further information and support, Philips Healthcare should be contacted at 1-800-263-3342.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- USA, Canada, Japan, United Arab Emirates, Austria, Switzerland, Germany, Spain, France, Italy, and Norway.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑