Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

FDA Recall Z-1916-2024 · Posted 2024-05-28

Recall StatusOpen, Classified
Product CodeHXX
Date Initiated2024-04-25
Product Quantity261 units

Reason for Recall

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Firm Action

DePuy Synthes issued Urgent Medical Device Recall (Removal) Letter on 4/26/24. Letter states reason for recall, health risk and action to take: Examine your inventory to determine if you have the subject product lots and quarantine them immediately. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Sales Consultant or contact the customer support services at ra-dpyuscsspecial@ its.jnj.com to coordinate the return/credits of the subject product lots. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMDField- [email protected] within three (3) business days of receipt of this notification. Please include in the email subject: FA 2363331. 4. Please complete the attached Business Response Form even if you do not have the subject product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage, transport, store, stock, or use the subject products). 6. If any of the subject product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.

Root Cause

Process change control

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑