Terminated

Triathlon MIS 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-706, 6541-5-707, & 6541-5-708; Triathlon MIS Instruments; Non Sterile: Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The Triathlon AR 4:1 Cutting Blocks and the Triathlon MIS 4:1 Cutting Blocks are used to prepare the Femur (bone cuts) for the Triathlon femoral component.

FDA Recall Z-1917-2009 · Posted 2009-08-26

Recall StatusTerminated
Product CodeLXH
Date Initiated2007-02-12
Date Terminated2009-10-15
Product Quantity690 - both varieties

Reason for Recall

Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle.

Firm Action

Important Product Correction letters were sent to all branches/agencies on February 12, 2008 by Federal Express. The letter asked consignees to examine their inventory and hospital locations to identify the affected product. Customers were asked to perform a functional inspection by assembling the Triathlon Impactor/Extractor Handle to all of their MIS Cutting Blocks. If the cutting block does not assemble, customers are to list the affected lot codes on the Attached Product Accountability Form and fax a copy to (201) 831-6069. The letter states that product should not be returned until further instructions are received. Questions should be directed to (201) 831-5825.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution