Terminated

SOMATOM go.Up, Model No. 11061628 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.

FDA Recall Z-1919-2019 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2019-06-04
Date Terminated2023-03-29
Product Quantity69

Reason for Recall

The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.

Firm Action

On June 4, 2019, the firm notified customers of the recall via Urgent Medical Device Correction letters. Customers were informed that the firm has developed software update CT VA20A-SP2A to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also improve some areas during the start-up and shut-down processes, as well as providing other improvements related to the system, examination and reconstruction processes, and dual energy and intervention . The corrective action will be provided free of charge and will be distributed via Update Instruction CT026/19P. This update will be performed remotely. If you have any questions, please contact our service organization at 1- 800-888-7436. Customers were asked to promptly notify and instruct all staff at their organizations who need to be aware of this notice. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product advisory notice. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution in the states of AL CA CT FL GA IL IN KS KY LA ME MI MO MS NC ND NE NH NY OH OK SD TN TX VA and WI