Medtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Ventricular Catheter, Catalog No. 46916. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region.

FDA Recall Z-1923-2014 · Posted 2014-06-27

Recall StatusTerminated
Product CodeJXG
Date Initiated2014-06-09
Date Terminated2015-01-14
Product Quantity1,187 units

Reason for Recall

Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.

Firm Action

An Urgent Medical Device Recall letter, dated 6/9/14, was sent to all their customers who purchased the Medtronic Duet External Drainage and Monitoring Systems. The letter informs the customers of the problem identified and the actions to be taken. Customers are instructed to cease use of the affected product, return any used units, account for any product used and verify that connections are in the condition as stated in the Instructions for Use for product currently in use. Additionally, customers are to complete the Customer Product Accountability Form. Customers with questions about the recall letter are instructed to contact Bob Shokoohi at (805) 571-8725.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- USA, Australia, Belgium, Canada, Croatia, Cyprus, Denmark, Egypt, Finland, France, Greece, Iran, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Switzerland, Turkey, and United Kingdom.