HDL-Cholesterol gen.3. For the quantitative determination of high-density lipoprotein Cholesterol (HDL-C) in serum and plasma.
FDA Recall Z-1924-2015 · Posted 2015-06-27
Advertisement
Recalling FirmRoche Diagnostics Operations, Inc.
Recall StatusTerminated
Product CodeLBS
Date Initiated2015-05-07
Date Terminated2017-03-24
Product Quantity6255 analytical units
Reason for Recall
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine
plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Firm Action
Roche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky
Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: [email protected]. For technical questions, contact the Customer Support Center at 1-800-428-2336
Root Cause
Under Investigation by firm
Distribution Pattern
US (nationwide) including PR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.