Dimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme Calibrator is an in vitro diagnostic product intended to be used to calibrate the Creatine Kinase MB Isozyme method for the Dimension(R) clinical chemistry system with heterogeneous immunoassay module.

FDA Recall Z-1925-2013 · Posted 2013-08-07

Recall StatusTerminated
Product CodeJIT
Date Initiated2013-06-05
Date Terminated2015-03-18
Product Quantity7,470

Reason for Recall

Siemens Healthcare Diagnostics has received complaints that the lyophilized cake is not fully dissolved after the stated time in the MMB calibrator Instructions For Use (IFU). The frequency of this occurrence is low but if it occurs, gel-like clumps may be observed.

Firm Action

The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated May 2013, to all consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed as follows: Until a permanent solution is identified, please follow the instructions listed in the letter in order to optimize calibration preparation; complete and return the attached Correction Effectiveness Check form via fax to the Siemens Technical Solutions Center at (302) 631-8467 and forward this notification to anyone to whom you may have distributed this product. If you have any questions, please contact the Siemens Technical Solutions Center or your local Siemens technical representative or call 800-441-9250.

Root Cause

Material/Component Contamination

Distribution Pattern

Nationwide distribution: US including states of : AK, AL, AR, AZ, CA, CO, CT, DC, GA, GU, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.