Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

FDA Recall Z-1926-2017 · Posted 2017-05-18

Recall StatusTerminated
Product CodeDOF
Date Initiated2014-11-29
Date Terminated2022-02-26
Product Quantity1016

Reason for Recall

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Firm Action

Magellan issued a follow-on Recall notification related to the FDA Safety Notification dated May 17, 2017. This current recall follows FDAs recommendation to remove venous blood samples as an allowable sample type from all LeadCare products. This recall will supersede and replace the prior Recall notifications. Notifications issued by : UPS Letter (US), Registered Letter, E-mail, or Phone calls form May 23-May 25. Questions: Reba Daoust, Director of QA/RA 978-248-4811 or [email protected] Refer to Firm's website: http://www.leadcare2.com/getmedia/2fd8a90f-5c64-4058-b6ab-5db13512f0fe/Letter-to-our-customers-5-17-17.pdf.aspx. The notification of May 2017 supercedes the firms previous communications:. Magellan Diagnostics, notified customers by letter dated Nov 29, 2014 and provided with the problem, mitigation action( customers have been advised to implement a minimum 24-hour incubation of the blood-treatment reagent mixture prior to analysis) . Customers were instructed to return a completed FaxBack form via fax to 978-600-1480. A revised letter issued May 1, 2017 recommending that customers make the following workflow changes: 1. Discontinue collecting blood samples for lead testing in Becton Dickinson (BD) EDTA Vacutainer tubes: Lavender and Tan tops. 2. Discontinue the 24 hour incubation for sample preparation. On 5/17/17 FDA issued News Release - FDA warns Americans about risk of inaccurate results from certain lead tests- manufactured by Magellan Diagnostics may provide inaccurate results for some children and adults. Blood drawn from a vein may provide results that are lower than the actual level in the blood. FDA is warning laboratories not to use Magellan Diagnostic lead tests wit blood drawn from the vein. At this time all Leadcare systems can be used with blood drawn from a finger or heel stick. Based on FDA/s warning the firm issued notification letter to users on 5/xx/17.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑