CLEO Patient Monitor /Tranquility VS Patient Monitor

FDA Recall Z-1926-2019 · Posted None

Recalling FirmInfinium Medical, Inc
Recall StatusOpen, Classified
Product CodeMWI
Date Initiated2019-05-31
Product Quantity95 devices

Reason for Recall

The CLEO & Tranquility VS Patient Monitors 510(k) indications for use in the United States did not include the temperature monitor accessory option.

Firm Action

Infinium Medical, Inc. notified customers on about 05/31/2019 via "URGENT: MEDICAL DEVICE RECALL" letter sent through email and certified letter. Instructions included to remove all temperature modules from the CLEO & Tranquility VS Patient Monitors and destroyed. They were also requested to complete and return the Medical Device Recall Return Response Form to [email protected]. Once the response form is received, a new IFU and standalone temperature monitor will be provided. Questions and comments can be sent to Mr. George Anello during regular business hours (8:00 AM - 04:30 PM EST) at QARAinfiniummedical.com or by phone at l-866-918-8434 ext. 109.

Root Cause

Labeling False and Misleading

Distribution Pattern

Distributed nationwide to California, Georgia, Louisiana, Florida, Maryland, Minnesota, Puerto Rico, South Carolina, Virginia, Wisconsin.