reSET-O Mobile App, iOS and Android, provides cognitive behavioral therapy, as part of substance abuse treatment.

FDA Recall Z-1926-2021 · Posted None

Recall StatusTerminated
Product CodePWE
Date Initiated2021-06-07
Date Terminated2023-06-14
Product Quantity3,370 units

Reason for Recall

Due to a software issue, patients with a urine drug screen (UDS) received access to a contingency management wheel spin for potential rewards regardless of the UDS results. Positive UDS results should not have resulted in access to a wheel spin.

Firm Action

On 05/20/2021, the Recalling Firm communicated via email an initial "URGENT: MEDICAL DEVICE RECALL" Notification with an amended/follow-up Notification Letter sent on or about 06/07/2021 informing customers that between 08/10/2020 and 04/30/2021, patients with a urine drug screen (UDS) logged into the Recalling Firm's database received access to a contingency management wheel spin for potential rewards regardless of the UDS results. Positive UDS results should not have resulted in access to a wheel spin. Customers are also being informed that there are no actions required for the safe handling of the corrected product. A software release was issued on 04/30/2021 to return the products to their intended operation. Given the potential for reduced clinical effectiveness, Healthcare Providers are encouraged to review patients that have a positive urine drug screen logged to ensure that the patient receives any treatment or counseling necessary to support the patient's treatment. Customers are asked to complete and return the "REPLY FORM" via Fax at 415-458-2700 (Attn: Yuri Maricich) or email to [email protected]. For questions or further assistance, call 1-833-697-3738 (available weekdays 8:00 a.m. - 8:00 p.m. EST)

Root Cause

Software design

Distribution Pattern

U.S.: AZ, CA, DE, FL, IL, IN, KY, MA, MD, MN, NC, NY, OH, TN, WA, and WI O.U.S.: N/A