Invue Split Driver, Assembly with Handle, Model SI70006 Rev G. Used in conjunction with the Invue Anterior Cervical Plate System under 501(k) clearance number K121060. The Invue Driver is an instrument consisting of a Radel handle, a shaft and hexalobe-like tip with a 4.0mm split, which allows the driver to collapse to fit into a screw and expand to retain the screw.

FDA Recall Z-1927-2013 · Posted 2013-08-08

Recalling FirmSpineFrontier, Inc.
Recall StatusTerminated
Product CodeHXX
Date Initiated2013-05-10
Date Terminated2015-07-20
Product Quantity12 devices

Reason for Recall

Device design.

Firm Action

SpineFrontier sent an URGENT Advisory Notice dated June 27, 2013, to all affected consignees regarding the use of Invue Drivers. The notice advises users that the Split Tip Stylet Driver Assembly (11-70099) and Sold Driver, AO Connection (11-70070 or 11-70077) are being shipped in place of the Split Driver Assembly w/Handle (SI70006). The uses of the replacement parts are explained. Consignees are instructed to complete the Acknowledgement Form and return it to SpineFrontier. On June 28, 2013, customers were sent a second Advisory Notice requesting the removal of the Split Driver Assy w/ Handle Part Number SI70006 Lot Numbers 027104 and 027105. Consignees were instructed to complete the Acknowledgement Form and return it to SpineFrontier. Consignees with questions were instructed to call 978-232-3990.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution including Texas, Ohio, and Missouri.