Recall StatusTerminated
Date Initiated2014-06-02
Date Terminated2015-02-05
Product Quantity3,839 kits
Reason for Recall
Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles
Firm Action
Siemens Healthcare Diagnostics issued Urgent Medical Device Recall (UMDR) to customers in the United States on June 02, 2014 and a Urgent Field Safety Notice (UFSN) issued outside the United States on June 02, 2014.
These notices inform the customer of the potential for swollen acid bottles due to confirmed metal contamination. Customers with alternative lots in inventory were instructed to transition to the alternative lot upon receipt of the communication. Customers without an alternative lot of acid in inventory were informed that Siemens proactively shipped an alternative lot to their laboratory and they should transition to that lot upon its receipt. Siemens also informed customers without an alternative lot that they may continue to run and report patient results using the affected lot of acid.
Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Root Cause
Material/Component Contamination
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of Canada, Argentina, China, Colombia, Korea, Malyasia, New Zealand, Mexico, Guatemala, and Singapore.