Terminated

R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 62 MM OD, QTY: (1), STERILE H2O2, REF 71335762. Orthopaedic.

FDA Recall Z-1928-2013 · Posted 2013-08-08

Recalling FirmSmith & Nephew Inc
Recall StatusTerminated
Product CodeMBL
Date Initiated2013-05-29
Date Terminated2016-03-22
Product Quantity53 units

Reason for Recall

One batch of 60 mm 20o R3 XLPE liner may be laser etched and packaged as a 62 mm 20o liner and one batch of 62 mm 20o R3 XLPE liner may be laser etched and packaged as a 60 mm 20o liner.

Firm Action

All affected Smith & Nephew Sales Representatives were notified of problem and the recall via email on 5/29. They were instructed to cease use and distribution of the product and quarantine for return. Surgeons were notified by letter beginning on 6/12. International consignees were notified on 05/29 via email.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution-USA including the states of PA, WA, TX, FL, IA, MI, WI, TN, KY, IN, NY, MO and OR, and the countries of India, Canada, Chile, Venezuela, Dubai, South Africa, Australia, Poland, Romania, Belgium, Germany and United Kingdom.