Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.

FDA Recall Z-1928-2020 · Posted 2020-05-08

Recall StatusTerminated
Product CodeDQK
Date Initiated2019-10-07
Date Terminated2021-04-27
Product Quantity35

Reason for Recall

Intermittent communication due to inconsistent crimping of connectors onto cables.

Firm Action

On xx/xx/2019, and URGENT MEDICAL DEVICE RECALL-ACTION REQUIRED, letter was sent to customers by FedEx. Customers are instructed to acknowledge this notification , contact the firm at the contact information in the communication to obtain a Returned Goods Authorization (RGA) number and to return all product to recalling firm.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, CA, FL, IL, IN and MO, and the countries of Germany, Netherlands, United Kingdom.