Terminated

MAS Omni Immune Controls

FDA Recall Z-1929-2014 · Posted 2014-06-26

Recall StatusTerminated
Product CodeJJY
Date Initiated2013-12-02
Date Terminated2014-07-08
Product Quantity3185

Reason for Recall

The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.

Firm Action

Urgent Medical Device Correction letters were sent on December 2, 2013 to all affected customers and distributors via trackable mail.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, China, Germany, Hong Kong and Malaysia.