EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.

FDA Recall Z-1929-2020 · Posted None

Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodeLPH
Date Initiated2019-10-22
Product Quantity86 units

Reason for Recall

The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.

Firm Action

The firm requested its agents to identify and return affected devices.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.