Terminated

Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended for the quantitative determination of CRP in human serum and plasma (lithium heparin). Measurement of C-Reactive Protein is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.

FDA Recall Z-1930-2009 · Posted 2009-09-08

Recall StatusTerminated
Product CodeDCN
Date Initiated2008-04-15
Date Terminated2009-09-09
Product Quantity25,760 cartons

Reason for Recall

falsely elevated results in EDTA plasma samples

Firm Action

Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice on April 15, 2008 informing their customers that the EDTA plasma specimen claim had been rescinded. Current lots in the firm's inventory have been put on hold and reworked with an Alert Card rescinding the EDTA plasma claim. Subsequent lots use a revised instruction for use stating "Do not use EDTA plasma" in the Specimen Collection and Handling Section. For further questions, contact Siemens Healthcare Diagnostics at 1-302-631-6564.

Root Cause

Labeling False and Misleading

Distribution Pattern

Worldwide Distribution -- US, Austria, Bosnia, Czech Republic, Hungary, Slovakia, Serbia, Denmark, Estonia, Latvia, Lithuania, Norway, Sweden, Belgium, Cyprus, Canary Islands, France, Great Britain, Germany, Ireland, Italy, Netherlands, Poland, Russia, Portugal, Saudi Arabia, Spain, South Africa, Switzerland, Turkey, Yemen, Bahamas, Brazil, Canada, Mexico, Australia, Bangladesh, China, Hong Kong, Japan, Malaysia, Pakistan, and Taiwan.