Terminated

VITROS 250 Chemistry System, Product Code 8132086, (UDI # 10758750004409) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.

FDA Recall Z-1933-2019 · Posted None

Recall StatusTerminated
Product CodeJJE
Date Initiated2019-04-08
Date Terminated2020-11-12
Product Quantity5020 units

Reason for Recall

Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.

Firm Action

On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions: - As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits. - If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides. - If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use. - Consider loading one cartridge of VITROS CK Slides on board your system at a time. - Complete and return the request form via fax to 1-888-557-3759. Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take.

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution