RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator.

FDA Recall Z-1933-2024 · Posted 2024-05-30

Recall StatusOpen, Classified
Product CodeQVA
Date Initiated2024-04-16
Product Quantity7 systems

Reason for Recall

A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.

Firm Action

On 04/29/2024, the firm sent an "Urgent: Medical Device Recall Safety Notification" via email to customer informing them that during internal dosimetric testing of SCINTIX OOS partials in periodic tracking mode, it was observed that the delivered dose error of the full fraction can exceed 5% with respect to the upper bound of one or more static volumes within the DVH in the plan (non-tracked, non-target volumes in the patient frame of reference, i.e., the planning CT). This was determined to be the result of a software defect associated with completing interrupted. Customers are instructed to: 1. Notify all clinical staff of this recall 2. Utilized the Clinical Guidance Document that is include with this notification. For questions or concerns - contact Service and Support at 650-239-9070 or email [email protected]

Root Cause

Software design

Distribution Pattern

U.S.: CA, CT, NJ, OR, PA and TX No O.U.S.

Official Source

View FDA Recall →