S4 Cervical (S4C) Revision Instruments (FW099R and FW101R). The S4C Revision Instruments assist in the removal of an occipital cervical screw during revision surgery.
FDA Recall Z-1934-2012 · Posted 2012-07-03
Advertisement
Recalling FirmAesculap Implant Systems LLC
Recall StatusTerminated
Product CodeHWW
Date Initiated2012-03-07
Date Terminated2012-08-08
Product Quantity15
Reason for Recall
Aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Instrument that is part of the S4C Occipital Instrument Set. The instruments may not thread into the screw properly during a revision surgery due to tolerance issues between the two instruments.
Firm Action
The firm initiated their recall of this product by issuing an "Important Product Removal Notification" letter to all consignees dated March 7, 2012. The letter described the problem, advised them not to use the product, and instructed consignees to return the instruments to the recalling firm with the attached Distribution Inventory Sheet. Customer Service may be contacted if questions at 1-866-229-3002.
Root Cause
Component design/selection
Distribution Pattern
Nationwide distribution, including the states of CA, CO, IL, LA, NV, OH, and TX .
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.