VITROS 250AT Chemistry System, Product Code 1758143, (UDI # 10758750000036) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
FDA Recall Z-1934-2019 · Posted None
Advertisement
Recalling FirmOrtho-Clinical Diagnostics
Recall StatusTerminated
Product CodeJJE
Date Initiated2019-04-08
Date Terminated2020-11-12
Product Quantity57 units
Reason for Recall
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
Firm Action
On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions:
- As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits.
- If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides.
- If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use.
- Consider loading one cartridge of VITROS CK Slides on board your system at a time.
- Complete and return the request form via fax to 1-888-557-3759.
Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take.
Root Cause
Device Design
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.