MEDLINE O.R. Necessities(TM) Sterile* Needle Counter, 40 Ct. Double Magnet / Foam Block with Blade Remover, reorder: NC100FBR Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.

FDA Recall Z-1935-2018 · Posted None

Recall StatusTerminated
Product CodeGAB
Date Initiated2018-04-24
Date Terminated2020-08-04
Product Quantity23150 units

Reason for Recall

The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.

Firm Action

Medline sent an Urgent Recall letter dated April 24, 2018. The letter identified the affected product, problem and actions to be taken. Customer were instructed to check inventory, quarantine all affected product and return completed enclosed response form. Those consignees who further distributed the product were directed to notify their customers. For question contact 866-359-1704.

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of: Canada, Saudi Arabia, Cayman Island, and United Arab Emirates