Open, Classified

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

FDA Recall Z-1935-2026 · Posted 2026-05-01

Recalling FirmCook Incorporated
Recall StatusOpen, Classified
Product CodeDQO
Date Initiated2026-04-02
Product Quantity271

Reason for Recall

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Firm Action

On April 2, 2026, the firm began notifying affected consignees via letters titled "URGENT: MEDICAL DEVICE RECALL." Customers were instructed to quarantine affected products in inventory and to immediately cease further distribution and use of the affected products. They should return the affected products to Cook Medical with a copy of the completed Acknowledgement and Receipt form to receive a product credit.

Root Cause

Under Investigation by firm

Distribution Pattern

US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →