Stirling Ultracold Portable Shuttle, ULT-25. Storage for cold-chain management, short-term or long-term ultra-low and storage and transport of biological specimens.

FDA Recall Z-1940-2014 · Posted 2014-07-01

Recalling FirmGlobal Cooling Inc
Recall StatusTerminated
Product CodeJRM
Date Initiated2014-03-20
Date Terminated2015-07-01
Product Quantity413 units

Reason for Recall

The external power supply input is overheating at the connection to the freezer.

Firm Action

Consignees received an Urgent Medical Device Recall letter, dated March 20, 2014, a Recall Response letter, a User Alert, and a label to attach to the freezer power supply input. The letter identified the reason for the recall and how to recognize that the device may fail. Customers are to discontinue use of the device if there are any signs of a loose connection, corrosion, discoloration, or melting; and take the necessary actions identified, including contacting Global Cooling Customer Service at [email protected]. The Recall Return Response was to be completed and returned.

Root Cause

Use error

Distribution Pattern

Worldwide Distribution -- USA, including the states of AK, AR, AZ, CA, CO, CT, FL, GA, GU, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Australia, Brazil, Canada, China, India, Indonesia, Japan, Korea, Kuwait, Mexico, Netherlands, Pakistan, Singapore, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.