Terminated

STERIS 4085, 5085 & 5085SRT Surgical Tables

FDA Recall Z-1941-2010 · Posted 2010-06-30

Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeFQO
Date Initiated2010-05-14
Date Terminated2012-02-24
Product Quantity79 units

Reason for Recall

STERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table

Firm Action

On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution: USA, Australia, India, Korea & New Zealand