Terminated

Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.

FDA Recall Z-1941-2012 · Posted 2012-07-03

Recall StatusTerminated
Product CodeINI
Date Initiated2010-11-26
Date Terminated2013-03-22
Product Quantity186

Reason for Recall

The main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.

Firm Action

Three (3) pieces of information was distributed in the notification. They are: (1) Letter of Correction to the Provider (2) Letter of Correction for the Consumer (provider to issue to consumer) (3) Product Insert for Victory 9 Controller Interface Harness Replacement. The letter of Correction for Victory 9 with Power Sear (SC609PS), dated November 2010, identified the affected product and the reason for the correction. Customers were to record the serial number(s) and contact their provider's office to make an appointment to have the harness replaced.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution