Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.
FDA Recall Z-1941-2012 · Posted 2012-07-03
Advertisement
Recalling FirmPride Mobility Products Corp
Recall StatusTerminated
Product CodeINI
Date Initiated2010-11-26
Date Terminated2013-03-22
Product Quantity186
Reason for Recall
The main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.
Firm Action
Three (3) pieces of information was distributed in the notification. They are: (1) Letter of Correction to the Provider (2) Letter of Correction for the Consumer (provider to issue to consumer) (3) Product Insert for Victory 9 Controller Interface Harness Replacement. The letter of Correction for Victory 9 with Power Sear (SC609PS), dated November 2010, identified the affected product and the reason for the correction. Customers were to record the serial number(s) and contact their provider's office to make an appointment to have the harness replaced.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.