illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.

FDA Recall Z-1941-2014 · Posted 2014-06-27

Recall StatusTerminated
Product CodeOMN
Date Initiated2014-04-22
Date Terminated2014-06-30
Product Quantity35 instruments

Reason for Recall

Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.

Firm Action

Meridian sent an Urgent Medical Device Recall letter dated April 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return their unit to Meridian Bioscience. Meridian's Technical Support Staff will call them directly to provide them with a return authorization number and instructions for decontaminating, packaging and shipping the unit. For additional information customers were instructed to call 800-343-3858 (North America or 513-271-3700 (Worldwide).

Root Cause

Employee error