Extractor Pro RX Retrieval Balloon Catheter 9-12 mm Above Retrieval Balloon Catheter; Universal Product Number (UPN) M00547000 The Extractor Pro RX Retrieval Balloon Catheters are intended for use endoscopically to remove stones from the biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon.
FDA Recall Z-1941-2015 · Posted 2015-07-01
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusTerminated
Product CodeFGE
Date Initiated2015-04-27
Date Terminated2016-01-27
Product Quantity39 units
Reason for Recall
One lot of the Extractor" Pro RX Retrieval Balloon Catheter (9-12mm balloon) contains an incorrect syringe size (12-15mm), which is larger than the size identified on the label.
Firm Action
Boston Scientific sent an Urgent Medical Device Recall Removal letter dated April 27, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify any affected product in their inventory, quarantine the affected product immediately and return to Boston Scientific in accordance with the enclosed recall instructions. Customers will receive replacements for all recalled product that is returned to Boston Scientific. Customers were instructed to forward the recall notification to their customers. Customers with questions were instructed to contact their local sales representative.
For questons regarding this recall call 763-494-1133.
Root Cause
Employee error
Distribution Pattern
Worldwide Distribution including the following countries: France, Great Britain, Spain, Ireland. .
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.