EVA Floor Lifts, Model Numbers: 400, 450, 600
FDA Recall Z-1943-2020 · Posted None
Advertisement
Recalling FirmHandicare AB
Recall StatusOpen, Classified
Product CodeFSA
Date Initiated2016-06-15
Product Quantity997 units
Reason for Recall
After feedback from customers and internal suspicion, Handicare has become aware that abnormal use of Eva-lifts could lead to a deterioation in the joint of the liftarm. This could lead to injury if the joint breaks. All lifts must therefore be checked and reported back to the manuafacturer.
Firm Action
The firm notified their consignees on 06/15/2016 by email. The notice explained the issue of wear in the joint of the lift arm due to abnormal use in heavy loads and/or many lifts. The consignee was directed to inspect the lifts in order to identify gaps and/or cracks and to report to the local Handicare sales representative and to Handicare.
On April 4, 2019, the firm sent a reminder notice on 04/04/2019
Root Cause
Device Design
Distribution Pattern
US and Mexico
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.