Margron DTC Hip Replacement System; Margron hip - Femoral stem 1+0, sterile, Portland Orthopaedics Inc., St. Clair, MI; REF 1-647-000. Product is used for Orthopedic surgery.
FDA Recall Z-1945-2008 · Posted 2008-09-18
Advertisement
Recalling FirmPortland Orthopaedics Pty, Ltd.
Recall StatusTerminated
Product CodeMEH
Date Initiated2008-01-11
Date Terminated2009-11-12
Product Quantity2403 of all products
Reason for Recall
Australian data indicates the product is associated with a higher than average rate of hip replacement revision surgery.
Firm Action
Stocks were pulled from U.S. consignees on or before 1/11/08. Consignee and implanting surgeon letters dated 6/20/08 were sent advising of the issue and recommending patient monitoring. For assistance, contact Portland Orthopaedics at 1-810-329-3348.
Root Cause
Device Design
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.