Tabs Professional Monitor Without Wall Bracket, Model 25023. Stanley Senior Technologies, Lincoln, NE.

FDA Recall Z-1945-2010 · Posted 2010-07-01

Recall StatusTerminated
Product CodeKMI
Date Initiated2010-03-26
Date Terminated2011-09-16
Product Quantity36 units

Reason for Recall

Product might not indicate low battery alarms or exit alarms when powered by a DC power supply.

Firm Action

The firm made phone calls beginning 3/26/10 to customers who purchased the 12-volt power supply informing them of the recall and instructing them to determine the number of units that will need to be returned for replacement. Important Safety Notice letters dated 3/25/10 were issued via regular mail on 3/26/10 and also providing the reason for the recall, instructions, and customer service contact information to arrange for return of the units.

Root Cause

Component change control

Distribution Pattern

Nationwide Distribution