ARJO Carendo Multi-Purpose Hygiene Chair; Model BIB2003-01, a non-AC powered patient lift; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden. The equipment is intended for hygiene care, showering, and toileting of hospital or nursing home residents under the supervision of trained nursing staff and in accordance with the instructions outlined in the Operating and Product Care Instructions.

FDA Recall Z-1946-2009 · Posted 2009-09-08

Recalling FirmArjoHuntleigh
Recall StatusTerminated
Product CodeFSA
Date Initiated2009-06-11
Date Terminated2010-08-19
Product Quantity1,987 hygiene chairs

Reason for Recall

Male patient genitals may be pinched between the seat and the frame of the patient lift chair when the Care Raiser seat is returned to its lowest position.

Firm Action

ArjoHuntleigh sent Urgent Device Field Correction letters dated June 11, 2009, along with the Urgent Field Safety Notice ID 090526AHE dated May 26, 2009, via UPS Ground delivery. Customers were informed of the affected device and were requested to complete the response form with new contact information and fax it back to ArjoHuntleigh at 630-567-5012 or 630-925-7969. For further questions, contact ArjoHuntleigh at 1-800-323-1245.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution