Terminated

Syva EMIT 2000 Theophylline Product Usage: The Syva Emit 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.

FDA Recall Z-1946-2018 · Posted None

Recall StatusTerminated
Product CodeKLS
Date Initiated2017-11-27
Date Terminated2019-04-19

Reason for Recall

Lot J1 of the Syva EMIT 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.

Firm Action

Siemens sent an Urgent Medical Device Recall letter dated November 27, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of affected product, review inventory to determine replacement needs, complete and return Field Correction Effectiveness Check list. For questions contact Siemens Customer Care Center or your local Siemens technical support representative.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide Distribution