PowerGlide ST Midline Catheter Maximal Barrier Kit-18 gauge,10 cm length Catalog Number: ST018101D
FDA Recall Z-1946-2020 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodePND
Date Initiated2020-03-17
Date Terminated2020-10-21
Product Quantity655 kits
Reason for Recall
Iincorrectly packaged without the required lidocaine as labeled
Firm Action
BD issued Urgent Medical Device Recall letter issued on March 17, 2020 via Fed'x or email states reason for recall, health risk and action to take:
1.Examine your inventory of PowerGlide ST" Midline Catheter Maximal Barrier for product code ST018101D, Lot No. REDW0476. Inventory of this kit lot may be used by the clinician through alternate sourcing of lidocaine prior to initiating the procedure. Alternatively, any remaining inventory may be discarded. BD will provide you with a replacement for each kit discarded.
2.Share this recall notification with all users of the product within your facility to ensure they are also aware of this recall.
3.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement.
4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse
If you require further assistance, please contact:
BD Contact Contact Information
Customer/Technical Support Phone: 800-290-1689
Monday Friday between 8:00 am to 5:00 pm MST in the United States
Root Cause
Packaging change control
Distribution Pattern
AZ, CT, GA, IL, LA, MA, MI, MN, ND, NE, NY, OH, SD, TX),