Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
FDA Recall Z-1948-2026 · Posted 2026-04-24
Advertisement
Recalling FirmMedtronic, Inc.
Recall StatusOpen, Classified
Product CodeOAD
Date Initiated2026-03-11
Product Quantity120698 units
Reason for Recall
Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
Firm Action
Medtronic issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 3/18/2026 via USPS and email. The notice explained the issue and investigations and provided procedure recommendations: Arrhythmia and or VT/ VF is a known potential adverse event during a cardiac ablation procedure, as indicated within the Sphere-9 IFU. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
As described in the Medtronic Sphere-9 IFU, defibrillation equipment must be available for immediate use in the case of a life-threatening arrhythmia. Further, the IFU recommends to not deliver energy when in contact with pacing and defibrillation leads.
CUSTOMER ACTIONS:
Follow the procedure recommendations listed in this letter in addition to the Instructions for Use (IFU) when using a Sphere-9 catheter with a concomitant Biotronik ICD/CRT-D.
Complete the enclosed Customer Confirmation Form and email to [email protected]. This form must be returned even if you do not have any affected product in your possession.
Please forward this notice to all those who need to be aware within your organization, including, but not limited to, physicians using the Sphere-9 catheter, and maintain a copy for your records.
*** Update: On 6/19/2026, Medtronic updated the field corrective action plan, including additional GTINs within scope, and a revised U.S. IFU
Root Cause
Device Design
Distribution Pattern
Worldwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.