Coseal Surgical Sealant 4mL, Product Codes: 934071 & 934074. Indicated for use in vascular reconstructions to achieve adjunctive hemostasis.
FDA Recall Z-1949-2010 · Posted 2010-07-02
Advertisement
Recalling FirmBaxter Bioscience
Recall StatusTerminated
Product CodeNBE
Date Initiated2010-01-14
Date Terminated2010-12-22
Product Quantity15,262
Reason for Recall
The recall is being conducted as a precautionary measure due to out of specification results at the 18 month and 21 month time points (out of the 24 month licensed shelf life to expiration) for the percent thiol substitution during a Coseal Stability study. The parameter that is out of specification is an indicator of possible failure for the product to gel appropriately.
Firm Action
The recall notification was initiated on January 14, 2010 with the firm sending a Product Recall letter and attached Customer Reply Form to the affected consignees. The recall letter informed the consignees of the reason for recall and requested them to immediately stop the usage/distribution of the lots noted above and to quarantine any units you may still have in your inventory. The letter provided instruction for product return.
The letter also requested the consignees to please forward the recall information as appropriate to other services/facilities/departments that they may have provided this product to. Consignees were instructed to complete the
Customer Reply Form, and fax it as soon as possible to 888-345-4261. Consignees with questions were instructed to contact Stericycle at 877-870-4486.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement