Respire Pink+ Hard Oral Sleep Apnea Device
FDA Recall Z-1949-2018 · Posted 2018-05-29
Advertisement
Recalling FirmRespire Medical
Recall StatusTerminated
Product CodeLRK
Date Initiated2017-09-01
Date Terminated2019-12-09
Product Quantity25 units
Reason for Recall
Potential for device breakage during use
Firm Action
On or about September 6, 2017, Respire & Whole You distributed untitled notices to their customers. Respire & Whole You notified customers that they received reports of device breakage. Respire & Whole You offered to send replacements to all impacted customers. On May 23, 2018, Respire & Whole You issued URGENT MEDICAL DEVICE RECALL notices and response forms to their customers. Customers were advised to take the following actions:
- Immediately examine your inventory, discontinue use and quarantine product subject to recall.
- In addition, if you may have further distributed this product, please identify your customers and notify them of the product recall by providing them with a copy of the notice.
- Once Respire & Whole You receives your returned device, the firm will remake and ship you a completely new Pink+ appliance with the original Scheu arms at no charge.
Customers with questions may call (718) 643-7326, Monday Friday 7am to 7pm.
Root Cause
Device Design
Distribution Pattern
US nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.