Terminated

Coseal Surgical Sealant 8mL, Product Codes: 934072 & 934075. Indicated for use in vascular reconstructions to achieve adjunctive hemostasis.

FDA Recall Z-1950-2010 · Posted 2010-07-02

Recalling FirmBaxter Bioscience
Recall StatusTerminated
Product CodeNBE
Date Initiated2010-01-14
Date Terminated2010-12-22
Product Quantity7,409

Reason for Recall

The recall is being conducted as a precautionary measure due to out of specification results at the 18 month and 21 month time points (out of the 24 month licensed shelf life to expiration) for the percent thiol substitution during a Coseal Stability study. The parameter that is out of specification is an indicator of possible failure for the product to gel appropriately.

Firm Action

The recall notification was initiated on January 14, 2010 with the firm sending a Product Recall letter and attached Customer Reply Form to the affected consignees. The recall letter informed the consignees of the reason for recall and requested them to immediately stop the usage/distribution of the lots noted above and to quarantine any units you may still have in your inventory. The letter provided instruction for product return. The letter also requested the consignees to please forward the recall information as appropriate to other services/facilities/departments that they may have provided this product to. Consignees were instructed to complete the Customer Reply Form, and fax it as soon as possible to 888-345-4261. Consignees with questions were instructed to contact Stericycle at 877-870-4486.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution

Official Source

View FDA Recall →