Revolution CT scanners Product Usage: The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
FDA Recall Z-1951-2019 · Posted None
Advertisement
Recalling FirmGE Healthcare, LLC
Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2019-01-19
Product Quantity857 (US = 195: OUS = 662)
Reason for Recall
Additional low dose radiation exposure.
Firm Action
GE Healthcare will send a customer letter that provide safety instructions for immediate mitigation of the issue. The customer can continue to use the system with the recommended workaround instruction provided in the customer letter: All affected systems will be corrected with updated software and a User Manual Addendum which re-emphasizes existing pat ient center ing instructions. If you have any questions or concerns regarding this notification. For question contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Root Cause
Device Design
Distribution Pattern
Worldwide - US Nationwide Distribution