Cios Alpha VA20/ VA30 mobile X-Ray systems
FDA Recall Z-1953-2019 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2019-06-20
Date Terminated2020-04-17
Product Quantity1
Reason for Recall
DAP chamber may be missing an insulating foil
Firm Action
1. Siemens will inspect the DAP chamber on your unit, and if necessary, replace it with insulated DAP chamber.
2. Siemens service organization will contact you shortly to arrange a date to perform this corrective action.
3. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Root Cause
Nonconforming Material/Component
Distribution Pattern
IL, NE
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.