Terminated

Stryker Reflex Hybrid Level Anterior Cervical Plate; Non Sterile; Manufactured by Stryker Spine SAS, Cestas, France. Distributed by Howmedica Osteonics Corp. Mahwah, NJ 07430.

FDA Recall Z-1958-2009 · Posted 2009-08-20

Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeKWQ
Date Initiated2006-06-14
Date Terminated2009-08-20
Product Quantity38987 plates

Reason for Recall

Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.

Firm Action

Stryker Spine issued an "Important Market Withdrawal" notice dated June 14, 2006. Consignees were advised to complete and return a Product Accountability Form via fax to 1-201-831-6069 prior to returning the affected product. Product return was to be performed during certain phase periods when upgrades were being shipped. For further questions, contact Stryker Spine at 1-201-831-5825.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution -- Including states of NC, OR, AZ, TN, CA, MA, MD, IL, GA, FL, NJ, NM, KY, LA, PA, MI, UT, MO, OK, KS, OH, MN, NE, IN, MT, TX, AL, CO and WI.