Recall StatusOpen, Classified
Date Initiated2025-05-23
Product Quantity4,882 units
Reason for Recall
Lack of sterility assurance.
Firm Action
A "1st NOTIFICATION - URGENT: VOLUNTARY MEDICAL DEVICE RECALL" notification dated 5/23/25 was mailed to consignees.
The notification instructs customers to identify affected devices in inventory and immediately place them in quarantine. The provided Acknowledgement Form is to be completed and returned via email to
[email protected] or FAX at 1-609-750-4220. Customers are to keep a copy of the notification for their records and forward a copy to users. A Returned Materials Authorization number and instructions to return product will be provided by Customer Service upon receipt of the completed Acknowledgement Form.
The notification instructs sales representatives and distributors to identify affected devices in inventory and immediately place them in quarantine. The provided Acknowledgement Form is to be completed and returned via email to
[email protected] or FAX at 1-609-750-4220. These individuals are instructed to keep a copy of the notification for their records and to forward the notice to individuals who received distributed product. Distributors are to collect response forms from their customers and provide them to Integra to obtain a Returned Materials Authorization number and directions to return product. Sales representatives will be contacted by Post Market Quality to provide recall instructions.
Additional communications were made to customers on 7/10/25 and 7/18/25 informing them that expired devices are also subject to this recall and should be disposed of.
Root Cause
Process change control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Taiwan (Province of China), Viet Nam, Brazil, Trinidad and Tobago, El Salvador, South Korea, Peru, Canada, Australia, New Zealand, Belgium, France, Germany, Ireland, Israel, Italy, Kenya, Slovenia, South Africa, Spain, Sweden, Switzerland and United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.