Remel X/pect Flu A&B Control Swabs, For In Vitro Diagnostic Use, Ref R246003, each package contains 20 Flu A Control Swabs and 20 Flu B Control Swabs, each swab is individually packaged in a foil pouch. The firm name on the label is Remel Europe, Ltd., Dartford, Kent, United Kingdom. Product is a test kit containing characterized positive and negative controls intended for use in qualitative procedures as external quality control material for the Remel Xpect Flu A&B test kit.
FDA Recall Z-1962-2013 · Posted 2013-08-13
Advertisement
Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeGNX
Date Initiated2013-02-19
Date Terminated2013-08-14
Product Quantity96 kits of 20 swab sets
Reason for Recall
Flu A+ Control Swabs incorrectly give Flu A- results when used for quality control testing with the Xpect Flu A&B test kit.
Firm Action
The firm, ThermoFisher Scientific, sent a "MEDICAL DEVICE RECALL" letter dated February 20, 2013 to its customers via regular mail. The letter described the product, problem and actions to be taken. The customers were instructed to review your inventory for affected product lots; discard the affected product; notify any personnel who needs to be notified of the potential for a false negative indication when using the swab, and complete and return the attached Medical Device Recall Return Response Acknowledgment & Receipt Form via Fax to: Attn: Technical Service & Regulatory Affairs at 1-877-428-1924.
If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).
Root Cause
Employee error
Distribution Pattern
Nationwide distribution: AL, AR, CA, CO, CT, FL, GA, IA, IL, KS, MA, MD, MI, MO, NM, NV, NY, OH, PA, TX, WA, and WI.